• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNBLNA
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2024
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer reported that the inpen was not logging the correct doses of insulin.Troubleshooting was performed and found that the intended dose amount and dose amount recorded in the inpen app was greater than 1.0 unit.It was found that the customer reported dose knob/dial was difficult to turn.No harm requiring medical intervention was reported.The customer will discontinue the use of the inpen and will be returned for analysis.
 
Manufacturer Narrative
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.Cartridge holder locks properly in place.Inpen paired with commercial mobile app with small dose.Received inpen [ 1/4 ] of travel.Performed encoder bond investigation and found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Also, plastic shavings contamination builds up found caused by encoder wheel tabs rubbing at the walls of the dose nut.Abrasions only at the top section of the dose nut.Unable to perform baseline/wireless functionality and displacement dose accuracy test.Performed leadscrew reset torque test.Inpen passed and is within specification (ccw: 3.901ozf-in)].Inpen passed front cap investigation.In conclusion: it was determined from destructive analysis that the difficult to dial/dose was caused by a pattern wheel misalignment due to an encoder base bond failure.This can affect insulin delivery.Unable to confirm inpen app not recording accurate doses due to pattern wheel misalignment.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNBLNA NOVO NORDISK BLUE
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key18718687
MDR Text Key335553277
Report Number3012822846-2024-00230
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/14/2023
Device Model NumberMMT-105NNBLNA
Device Catalogue NumberMMT-105NNBLNA
Device Lot NumberB0635
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2024
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received02/21/2024
Supplement Dates FDA Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-