| Model Number |
15-8522/57 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
01/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information received from surgeon (b)(6) 2024 suggests that the design of this cone and stem construct may not have been well suited for this patient: the surgeon noted on (b)(6) 2024 that the wl cones address metaphyseal area but do not address meta-diaphyseal ingrowth, especially when used with cementless stems (as in this case).The subject components were implanted by dr.(b)(6) during revision knee surgery on (b)(6) 2022.Patient complained of pain located in medial and lateral right knee starting after surgery.Due to pain and femoral stem loosening, the patient's knee was revised on (b)(6) 2024.Femoral side implants were replaced with custom-made femoral stem (lbio ref: cus-736, wl ref: (b)(4)), and standard items 15-8521/10 (total condylar replacement) and 15-0027/15 (replacement plateau).[customer].
|
| |
|
Additional Manufacturer Narrative
|
|
This is the final supplemental report, the complaint is closed.
|
| |
|
Event or Problem Description
|
|
Information received from surgeon (b)(6) 2024 suggests that the design of this cone and stem construct may not have been well suited for this patient: the surgeon noted on (b)(6) 2024 that the wl cones address metaphyseal area but do not address meta-diaphyseal ingrowth, especially when used with cementless stems (as in this case).The subject components were implanted by dr.M.During revision knee surgery on (b)(6) 2022.Patient complained of pain located in medial and lateral right knee starting after surgery.Due to pain and femoral stem loosening, the patient's knee was revised on (b)(6) 2024.Femoral side implants were replaced with custom-made femoral stem (lbio ref: (b)(6), wl ref: case 2023_0468), and standard items (b)(6) (total condylar replacement) and (b)(6) (replacement plateau).[customer].
|
| |
|
Additional Manufacturer Narrative
|
|
This is the final supplemental report, the complaint is closed.
|
| |
|
Event or Problem Description
|
|
Information received from surgeon 18-jan-2024 suggests that the design of this cone and stem construct may not have been well suited for this patient: the surgeon noted on (b)(6) 2024 that the wl cones address metaphyseal area but do not address meta-diaphyseal ingrowth, especially when used with cementless stems (as in this case).The subject components were implanted by dr.(b)(6) during revision knee surgery on (b)(6) 2022.Patient complained of pain located in medial and lateral right knee starting after surgery.Due to pain and femoral stem loosening, the patient's knee was revised on (b)(6) 2024.Femoral side implants were replaced with custom-made femoral stem (lbio ref: (b)(4), wl ref: case (b)(4)), and standard items 15-8521/10 (total condylar replacement) and 15-0027/15 (replacement plateau).[customer].
|
| |
|