| Model Number |
IPN915303 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "nasal plug missing on the device in an unopened packaging".The issue was identified in the warehouse.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Qn # (b)(4).The reported complaint of "additional/missing component - missing" was confirmed based upon the sample received.Visual examination of the returned device revealed that only the syringe was returned in the packaging.The packaging was confirmed to be sealed properly.A device history record review was performed on the device with no evidence to suggest a manufacturing related root cause.The mad device was missing.An investigation within teleflex has been initiated to further investigate this complaint issue.
|
| |
|
Event or Problem Description
|
|
It was reported that "nasal plug missing on the device in an unopened packaging".The issue was identified in the warehouse.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|