Model Number KD-V431M-0720
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During device evaluation at olympus, it was found there was biomaterial inside the distal tip indicating use.There were minor charred stains on the cutting wire and the blue coated portion was torn and peeled.The slide was working properly and the distal end could be bent smoothly to the desired direction and angle.The tome was checked with a test generator and cord using monopolar setting 120w.The cutting wire could generate energy output and stayed intact after multiple activation attempts.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause of the torn coating could not be determined, likely factors causing the tear could have been the following: 1.The slider was pushed more than needed causing the cutting wire to deflect.2.The deflected coated portion of the cutting wire and the metal part of the distal end of the endoscope came into contact when the forceps elevator was raised.3.The tube was moved back and forth in the situation of description ¿2¿, causing the metal part of the distal end of the endoscope to scratch the coated portion of the wire.As a result, the coated portion of the wire was ruptured.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu : - when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.- do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
The customer reported wire of the single use 3-lumen sphincterotome was broken.The cutting wire that enabled the tip of the device to bend was broken and the customer could not get the tip to bend when testing it.The issue occurred after opening the device package prior to a therapeutic procedure.There was no reported patient harm or impact due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18719848
Report Number9614641-2024-00396
Device Sequence Number13146235
Product Code KNS
UDI-Device Identifier04953170382628
UDI-Public04953170382628
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V431M-0720
Device Lot Number16V 01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Initial Report FDA Received Date02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1