During device evaluation at olympus, it was found there was biomaterial inside the distal tip indicating use.There were minor charred stains on the cutting wire and the blue coated portion was torn and peeled.The slide was working properly and the distal end could be bent smoothly to the desired direction and angle.The tome was checked with a test generator and cord using monopolar setting 120w.The cutting wire could generate energy output and stayed intact after multiple activation attempts.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause of the torn coating could not be determined, likely factors causing the tear could have been the following: 1.The slider was pushed more than needed causing the cutting wire to deflect.2.The deflected coated portion of the cutting wire and the metal part of the distal end of the endoscope came into contact when the forceps elevator was raised.3.The tube was moved back and forth in the situation of description ¿2¿, causing the metal part of the distal end of the endoscope to scratch the coated portion of the wire.As a result, the coated portion of the wire was ruptured.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu : - when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.- do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
|