It was reported that the patient underwent an initial left unicompartmental knee arthroplasty.Subsequently, after a twisting movement, the patient experienced pain with an onset of limited range of motion.Revision surgery was performed due to dislocation of the inlay.During the revision, the inlay was found fractured with a defect in the transition from the medial to dorsal underside and continuous on the lateral side.Additionally, the tibial component was slightly medialized.The femoral component remained in place and the tibial and inlay were revised with competitor product.
|
|
(b)(4).This follow-up report is being submitted to relay additional and corrected information.The following section was corrected: b3.The following sections were updated: b4, b5, d9, g3, g6, h2, h3, h6, h11.The reported bearing was returned and assessed.It exhibits wear as reported on both articulating faces and also exhibits a crack; however, the crack does not extend through the entire length of the product, so the bearing remains structurally intact.The bearing has not been submitted for fracture analysis because the process would require the bearing to be cut in order to have a clear view on the fracture surface.Additionally, there is uncertainty regarding the decontamination method used by the hospital.If an autoclave was employed, it may have altered the material structure, potentially compromising the reliability of any fracture analysis.Review of the complaint identified that the devices have been implanted for approximately 11 years.The associated tibial tray was also returned.There appears to be minimal bone cement attached however it is not clear if the cement has been chipped off to reveal the item identification as there are witness marks of something which might have been used to scrape the bone cement.The bearing surface of the tray exhibits scuffing in parallel to the rail between the bearing and tibial tray bearing surface.There is also a large area of damage to the edge of the tibial tray bearing surface.It is not clear if these may have been caused by the removal of the tray during the revision surgery.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and lot combination.Medical records were provided and reviewed by a health care professional.The review identified that the revision op notes state the following.¿ diagnosis: inlay subluxation.¿ change of the tibial component to cemented monoblock implant: medacta.¿ according to the patient, after a twist movement, pain with a recurrent feeling of blockage occurred.Up to that point, the patient had been satisfied.¿ ¿the tibial component appears slightly medialized in relation to the femur¿.¿ ¿extension of the inlay to the lateral boundary of the prosthesis shows a recurrence of the inlay, and a ventral osteophyte femorally limits the movement of the lnay ventrally by a mechanical collision.Removal of the inlay.This shows a defect in the transition from the medial to the dorsal third on the underside of the inlay.On the medial side, the defect is continuous, on the lateral side there is a narrow bar, so that the inlay is not completely broken in two.¿ ¿ no bony defect formation on patella.¿ anterior cruciate ligament is sparsley intact.¿ sagittal there is a slightly increased tibial mobility with a terminally fixed stop.¿ femoral- no damage to metallic surfaces and without signs of loosening.¿ new tibial component cemented.¿ no intra-op complication/events.Radiographs were provided and reviewed by a radiologist.The review identified that pre-revision; the tibial component is slightly medially positioned with respect to the femoral component.Sizing is appropriate.There is mild osteopenia but overall adequate bone quality.Polyethylene dislocation is noted on the study with suggestion of medial and ventral position.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|