On 14th february 2024, getinge became aware of an issue with one of our surgical lights ¿ volista standop 400/400.As it was stated, the water was present inside the device.There was no injury reported, however, we decided to report the issue in abundance of caution as contact of water with live parts may cause electric shock in case of reoccurrence.
|
|
On 14th february 2024, getinge became aware of an issue with one of our surgical lights ¿ volista standop 400/400.As it was stated, the water was present inside the device.As it was further confirmed with the photographic evidence, the traces of water/cleaning products were present on the inside of underside cover.There was no injury reported, however, we decided to report the issue in abundance of caution as contact of water with live parts may cause electric shock in case of reoccurrence.
|
|
Initial reporter: hospital medical equipment.The correction of h4 manufacture date deems required.This is based on the internal evaluation.Previous h4 manufacture date: 2019-11-15.Corrected h4 manufacture date: 2019-11-18.The correction of b5 describe event and problem deems required.This is based on the internal evaluation.Previous b5 describe event and problem: on 14th february 2024, getinge became aware of an issue with one of our surgical lights ¿ volista standop 400/400.As it was stated, the water was present inside the device.There was no injury reported, however, we decided to report the issue in abundance of caution as contact of water with live parts may cause electric shock in case of reoccurrence.Corrected b5 describe event and problem: on 14th february 2024, getinge became aware of an issue with one of our surgical lights ¿ volista standop 400/400.As it was stated, the water was present inside the device.As it was further confirmed with the photographic evidence, the traces of water/cleaning products were present on the inside of underside cover.There was no injury reported, however, we decided to report the issue in abundance of caution as contact of water with live parts may cause electric shock in case of reoccurrence.The correction of h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code - explain fields deems required.This is based on the internal evaluation.Previous h3a device evaluated by manufacturer: no.Corrected h3a device evaluated by manufacturer: yes.Previous h3b device not eval provide code: other.Corrected h3b device not eval provide code: n/a.Previous h3c if other provide code - explain: device not returned to manufacturer.Corrected h3c if other provide code - explain: n/a.The correction of d4 catalog # and unique identifier (udi) # deems required.This is based on the internal evaluation.Previous d4 catalog # ardvst229013a.Corrected d4 catalog # blank.Previous d4 unique identifier (udi) #: n/a.Corrected d4 unique identifier (udi) #: (b)(4).The correction of d1 brand name deems required.This is based on the internal evaluation.Previous d1 brand name volista standop.Corrected d1 brand name standop volista®.Getinge became aware of an issue with one of our surgical lights ¿ volista standop 400/400.As it was stated, the water was present inside the device.As it was further confirmed with the photographic evidence, the traces of water/cleaning products were present on the inside of underside cover.There was no injury reported, however, we decided to report the issue in abundance of caution as contact of water with live parts may cause electric shock in case of reoccurrence.It was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.There is no information if the claimed device was not being used for patient treatment or diagnosis when the event took place.It was reported the presence of residue coming from cleaning product inside the under face of the cupola.The root cause is a misuse during cleaning of the medical device (excessive cleaning leading to water ingress).It is possible that a deterioration of the underside may occur over time (e.G.: whitish spot, small streak, hazy aspect¿).The user manual mentions to check the lightheads for chipped paint, impact marks and any other damages during daily checks.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
|