• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEGADYNE MEDICAL PRODUCTS, INC. 10FT SMOKE PENCIL UCONN; RETURN ELECTRODE CABLE, DUAL PLATE Back to Search Results
Catalog Number 251010J
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the removal of infected hardware that while using coag on setting of 30/30 along with a metal forceps the bovie tip caught fire but did not spark.The tip was replaced and the procedure was completed with the same pencil and generator.There were no adverse consequences for the patient, surgeon, and staff.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/16/2024.No lot or batch number was provided therefore a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/22/2024 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the 251010j device was received with no apparent damage.The device was connected to the generator and it worked as intended.There were no anomalies noted with the functionality of the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the filter performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10FT SMOKE PENCIL UCONN
Common Device Name
RETURN ELECTRODE CABLE, DUAL PLATE
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
PR  
3035526892
MDR Report Key18720523
Report Number1721194-2024-00016
Device Sequence Number20230662
Product Code GEI
UDI-Device Identifier10614559105573
UDI-Public10614559105573
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K982130
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number251010J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2024
Supplement Date Received by Manufacturer02/16/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-