• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00534700
Medical Device Problem Codes Break (1069); Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to boston scientific corporation that an advanix-naviflex rx biliary stent was to be used in the biliary tree during an endoscopic retrograde cholangiopancreatography (ercp) - biliary plastic stent implant procedure performed on an unknown date.During the procedure, the guide catheter detached from the push catheter during attempted deployment.The detached guide catheter was removed using retrieval forceps and the procedure was completed using another device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of guide catheter break.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of guide catheter break.Imdrf impact code f2301 captures the reportable event of the guide catheter being removed using forceps.Block h10: the advanix-naviflex delivery system was received for analysis; the stent was not returned.Visual examination of the returned device found the guide catheter detached, the pull wire kinked, and the push catheter suture hole torn.No other damages were noted to the delivery system.The reported event of guide catheter break was confirmed.Taking all available information into consideration, the investigation concluded that the reported event and observed damages were likely due to factors encountered during the procedure.It is possible that tortuosity encountered during the procedure and/or the excessive force applied to pull the device, limited the performance of the device and contributed to the reported event and observed problems.Therefore, the most probable cause of the reported event is adverse event related to procedure.A labeling review was performed and, from the information available, there is no information that this device was used in a manner inconsistent with the instructions for use (ifu)/ product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix-naviflex rx biliary stent was to be used in the biliary tree during an endoscopic retrograde cholangiopancreatography (ercp) - biliary plastic stent implant procedure performed on an unknown date.During the procedure, the guide catheter detached from the push catheter during attempted deployment.The detached guide catheter was removed using retrieval forceps and the procedure was completed using another device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18720741
Report Number3005099803-2024-00428
Device Sequence Number20402245
Product Code FGE
UDI-Device Identifier08714729787259
UDI-Public08714729787259
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K101314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2025
Device Model NumberM00534700
Device Catalogue Number3470
Device Lot Number0031454834
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2024
Supplement Date Received by Manufacturer03/26/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-