| Model Number |
M00534700 |
| Medical Device Problem Codes |
Break (1069); Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an advanix-naviflex rx biliary stent was to be used in the biliary tree during an endoscopic retrograde cholangiopancreatography (ercp) - biliary plastic stent implant procedure performed on an unknown date.During the procedure, the guide catheter detached from the push catheter during attempted deployment.The detached guide catheter was removed using retrieval forceps and the procedure was completed using another device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of guide catheter break.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0401 captures the reportable event of guide catheter break.Imdrf impact code f2301 captures the reportable event of the guide catheter being removed using forceps.Block h10: the advanix-naviflex delivery system was received for analysis; the stent was not returned.Visual examination of the returned device found the guide catheter detached, the pull wire kinked, and the push catheter suture hole torn.No other damages were noted to the delivery system.The reported event of guide catheter break was confirmed.Taking all available information into consideration, the investigation concluded that the reported event and observed damages were likely due to factors encountered during the procedure.It is possible that tortuosity encountered during the procedure and/or the excessive force applied to pull the device, limited the performance of the device and contributed to the reported event and observed problems.Therefore, the most probable cause of the reported event is adverse event related to procedure.A labeling review was performed and, from the information available, there is no information that this device was used in a manner inconsistent with the instructions for use (ifu)/ product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an advanix-naviflex rx biliary stent was to be used in the biliary tree during an endoscopic retrograde cholangiopancreatography (ercp) - biliary plastic stent implant procedure performed on an unknown date.During the procedure, the guide catheter detached from the push catheter during attempted deployment.The detached guide catheter was removed using retrieval forceps and the procedure was completed using another device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|