• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS CHLORIDE ELECTRODE; ELECTRODE, ION-SPECIFIC, CHLORIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS CHLORIDE ELECTRODE; ELECTRODE, ION-SPECIFIC, CHLORIDE Back to Search Results
Catalog Number 03003523001
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the integra 400 plus analyzer is (b)(6).The field service engineer determined there was a poorly fitting sealing ring on the reference electrode, causing air to enter.The electrode had recently been replaced.The sealing rings were cleaned and checked for integrity; they were in acceptable condition.The cl electrode was changed.Diagnostic tests were acceptable.Maintenance was performed on the ise module.Calibration and controls were run and expected values were measured.The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with the chloride electrode on a cobas integra 400 plus.No questionable results were reported outside of the laboratory.The sample resulted in a cl value of 120 mmol/l when tested on the integra 400 plus analyzer.The sample was repeated on an unknown gasometer instrument, resulting in a cl value of 109 mmol/l.This value was considered to be correct by the customer.
 
Additional Manufacturer Narrative
The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORIDE ELECTRODE
Common Device Name
ELECTRODE, ION-SPECIFIC, CHLORIDE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18720798
Report Number1823260-2024-00460
Device Sequence Number5740843
Product Code CGZ
UDI-Device Identifier08430215015353
UDI-Public08430215015353
Combination Product (Y/N)Y
Initial Reporter CountryBR
PMA/510(K) Number
K060373
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03003523001
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/02/2024
Supplement Date Received by Manufacturer02/26/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date02/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-