• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION GLIDESHEATH SLENDER; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION GLIDESHEATH SLENDER; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A5: ethnicity: not provided for animal use.A6: race: not provided for animal use.D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.The actual device has not been returned for evaluation.The investigation is currently ongoing.A followup report will be submitted once the investigation is complete.
 
Event or Problem Description
The user facility reported that the hub was leaking from the back of the glidesheath slender sheath.The procedure type was radial access heart catheterization.The patient was in stable condition and there was no blood loss.The event occurred intra-operative.There was no patient injury medical or surgical intervention required.There is not a direct allegation that the reported device caused or contributed to patient injury and/or need for medical intervention.Additional information was received on 06 feb 2024: a glidewire was used with the glidesheath slender at the time of leaking.The leak came from the valve area.The leaking was not significant.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.As of 24 apr 2024, the sample has not been returned for assessment.Should the sample arrive, the complaint will be reopened.The complaint cannot be confirmed for valve leakage because the complaint sample was not returned for assessment.The exact root cause could not be determined.Historically, valve leakages have been caused by off-center insertion of the dilator.Review of the device history record (dhr) showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the design failure mode and effects analysis (dfmea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESHEATH SLENDER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
gina digioia
950 elkton blvd
elkton, MD 21921
4103927218
MDR Report Key18721973
Report Number1118880-2024-00016
Device Sequence Number6350261
Product Code DYB
UDI-Device Identifier00389701012117
UDI-Public00389701012117
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K173831
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2026
Device Model NumberN/A
Device Catalogue Number80-1060
Device Lot Number0000446428
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2024
Supplement Date Received by Manufacturer04/29/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date05/14/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLIDEWIRE
Patient SexUnknown
-
-