H.6 investigation summary this statement is to summarize the investigation of a complaint involving the bd kiestra inoqula.According to the information provided, the issue was reported that there were plates found with no growth.During the investigation, it was reported that after previous occurrences, a camera was installed to monitor the situation.After the camera was installed, the customer reported again an occurrence of the issue.The issue was escalated to the system support specialist (sss) and research and development (r&d) for further investigation.The photos of the negative plates were retrieved, were reviewed by r&d, and showed that the beads went through the plate.Despite the issue reoccurring, the customer transitioned from the old zeus to the zeus ii system and the issue did not reoccur after this change.Following this action, no further issues were encountered.Based on the investigation, this case has been assessed as confirmed for a bd quality issue.Review found no new trends, risks, or hazards were identified because of this complaint.The issue in this complaint does not require the initiation of a corrective and preventative action (capa).A design history record (dhr) review is not required for this complaint.The complaint was evaluated via other elements of the investigation.Bd quality will continue to closely monitor for trends associated with this issue.
|