EXACTECH INC. NV GXL LINR, NTRL, 28MM ID, GROUP 1 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
|
Back to Search Results |
|
| Catalog Number |
130-28-51 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
10/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10.Concomitants: 5199027, 190-10-02 - alt ha clr std sz 2.5292175, 170-28-03 - biolox delta femoral head 28mm od, +3.5mm.6334662, 186-01-48 - integrip cc, cluster 48mm, g1.S069531, 180-65-35 - alteon 6.5mm screw, 35mm.These devices are used for treatments.Not diagnosis.Pending investigation.There is no other information available.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a left total hip arthroplasty on (b)(6) 2020 and then experienced a revision surgical procedure on (b)(6) 2023 approximately 3 years and 3 months after initial implant.Revision op report stated that the polyethylene liner had significant catastrophic wear.No images of the devices are provided.The device will not be returned.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Investigation results - the most likely cause for the revision reported due to prosthesis wear is a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.However, this is not confirmed, the devices were not returned.These devices are used for treatments, not diagnosis.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
The most likely underlying cause for the revision due to early prosthesis wear reported is a combination of risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.".
|
| |
|