BECTON DICKINSON BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367281
Medical Device Problem Code Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd vacutainer® safety-lok¿ blood collection set, the wingset didn't gain suction and collect blood during use.One quantity affected.No patient impact or injury reported.
 
Additional Manufacturer Narrative
The manufacturing location for this product is nipro.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields have been updated with additional information.Customer provided clarifying information regarding event details.B5: it was reported that while using bd vacutainer® safety-lok¿ blood collection set, devices are leaking blood from the np needle when inserted into the vacuette tubes.One box was affected.No patient impact reported.D10: device available for evaluation: yes.D10: returned to manufacturer on: 08-mar-2024.H.6.Investigation summary: material #: 367281.Lot/batch #: 23e0411.Bd received 5 samples for investigation.The samples were evaluated by visual examination and functional leak testing and the indicated failure mode for leakage with the incident lot was not observed.Additionally, 50 retention samples from bd inventory were evaluated by visual examination and another 8 by functional leak testing and no issues were observed relating to leakage as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode leakage.Bd was not able to identify a root cause for the indicated failure mode.
 
Event or Problem Description
It was reported that while using bd vacutainer® safety-lok¿ blood collection set, devices are leaking blood from the np needle when inserted into the vacuette tubes.One box was affected.No patient impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18723662
Report Number2243072-2024-00184
Device Sequence Number5525409
Product Code JKA
UDI-Device Identifier00382903672813
UDI-Public(01)00382903672813
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K991088
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2026
Device Catalogue Number367281
Device Lot Number23E0411
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2024
Supplement Date Received by Manufacturer02/26/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date03/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-