| Model Number |
101/860/090 |
| Medical Device Problem Codes |
Disconnection (1171); Separation Problem (4043)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/29/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the pilot balloon became detached during patient use.The tracheostomy tube was inserted on september 15th.The patient was removed from ventilator on september 19th.Moved to a controlled oxygen mask.Nurse visited patient room on october 2nd and found the pilot balloon deflated and disconnected from inflation line.Doctor replaced the tube and light blown colored liquid (gastric fluid) could be aspirated.Patient injury was unknown.However accidental swallowing to the trachea possibly occurred on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: d3,g1, and g2 email is: regulatory.Responses@icumed.Com.Device evaluation: one device was returned for investigation.Under visual inspection we noticed that inflation line was detached from pilot balloon as reported by customer.The complaint was confirmed.What caused the failure of the supplied item could not be established.Dhr review not completed as the issue is not lot specific.The complaint issue continues to be monitored and suitable actions will be taken as appropriate.This issue will continue to be monitored and further actions taken accordingly.
|
| |
|