• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLU SELCT SUCTIONAID CUFFED NONFENESTRATED; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLU SELCT SUCTIONAID CUFFED NONFENESTRATED; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/090
Medical Device Problem Codes Disconnection (1171); Separation Problem (4043)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the pilot balloon became detached during patient use.The tracheostomy tube was inserted on september 15th.The patient was removed from ventilator on september 19th.Moved to a controlled oxygen mask.Nurse visited patient room on october 2nd and found the pilot balloon deflated and disconnected from inflation line.Doctor replaced the tube and light blown colored liquid (gastric fluid) could be aspirated.Patient injury was unknown.However accidental swallowing to the trachea possibly occurred on the patient.
 
Additional Manufacturer Narrative
Other text: d3,g1, and g2 email is: regulatory.Responses@icumed.Com.Device evaluation: one device was returned for investigation.Under visual inspection we noticed that inflation line was detached from pilot balloon as reported by customer.The complaint was confirmed.What caused the failure of the supplied item could not be established.Dhr review not completed as the issue is not lot specific.The complaint issue continues to be monitored and suitable actions will be taken as appropriate.This issue will continue to be monitored and further actions taken accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLU SELCT SUCTIONAID CUFFED NONFENESTRATED
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
reed covert
MDR Report Key18724827
Report Number3012307300-2024-10011
Device Sequence Number15040350
Product Code BTO
UDI-Device Identifier15019517076172
UDI-Public15019517076172
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number101/860/090
Device Catalogue Number101/860/090CZ
Device Lot Number4276630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2023
Initial Report FDA Received Date02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight58 KG
Patient EthnicityNon Hispanic
-
-