ROCHE DIAGNOSTICS COBAS U 411 URINE ANALYZER; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 04906969001
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The test strip lot number and expiration date were requested, but not provided.The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they received questionable results for an unspecified number of patient samples tested on a cobas u 411 urine analyzer.Multiple test parameters are affected.The customer provided an example of one patient sample with discrepant leukocyte values.The sample initially resulted in a leukocyte value of 500 leukocytes/ul.The same sample was repeated, resulting in a negative leukocyte value.
 
Additional Manufacturer Narrative
The quality controls failed.The field service engineer replaced the photometer and the instrument was working properly after this action.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS U 411 URINE ANALYZER
Common Device Name
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18725239
Report Number1823260-2024-00467
Device Sequence Number20230698
Product Code KQO
UDI-Device Identifier04015630925636
UDI-Public04015630925636
Combination Product (Y/N)Y
Initial Reporter CountryPL
PMA/510(K) Number
K093555
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04906969001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2024
Supplement Date Received by Manufacturer03/15/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date03/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-