| Catalog Number |
04906969001 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The test strip lot number and expiration date were requested, but not provided.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received questionable results for an unspecified number of patient samples tested on a cobas u 411 urine analyzer.Multiple test parameters are affected.The customer provided an example of one patient sample with discrepant leukocyte values.The sample initially resulted in a leukocyte value of 500 leukocytes/ul.The same sample was repeated, resulting in a negative leukocyte value.
|
| |
|
Additional Manufacturer Narrative
|
|
The quality controls failed.The field service engineer replaced the photometer and the instrument was working properly after this action.The investigation determined the service actions resolved the issue.
|
| |
|