| Catalog Number |
1011340-40 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a right internal carotid lesion.During deployment of an 8x40mm acculink stent, the stent jumped forward and did not cover the entire target lesion.An 8x30mm acculink stent was deployed to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.It is possible that as the stent was being deployed interaction with the anatomy resulted in preventing the shaft lumens from moving freely resulting in the stent to jump in a spring like release; thus resulting in the reported malposition of device (stent jumped forward and did not cover the entire lesion) however, this could not be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|