Catalog Number 18060150
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
As reported by rep: "strike plate snapped from the targeting adapter while inserting the nail (hitting the strike plate using the hammer)".
 
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned, and matches the alleged failure.The returned targeting arm and strike plate were visually inspected where visible scratch and nick marks are on the device which indicate usage over some time.The broken threads of the strike plate are stuck inside a threaded hole meant for the strike plate in the nail adapter, after breaking.The broken thread portion of the strike plate was not returned.The broken surface of the thread indicates a sign of a brittle fracture evident by abrupt breakage no definite fracture progression lines on the fractured surface and no presence of plastic deformation as well.The strike plate may have been not seated flush with the target device and impacted in a non-axial direction.The assembly hole of the targeting arm is deformed which indicates the strike plate was hit from the side.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to being user related.If any further information is provided, the complaint report will be updated.
 
Event or Problem Description
As reported by rep: "strike plate snapped from the targeting adapter while inserting the nail (hitting the strike plate using the hammer)".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRIKE PLATE T2 FEMUR
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18727691
Report Number0009610622-2024-00074
Device Sequence Number5741077
Product Code LXH
UDI-Device Identifier04546540196088
UDI-Public04546540196088
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number18060150
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2024
Supplement Date Received by Manufacturer04/09/2024
Initial Report FDA Received Date02/16/2024
Supplement Report FDA Received Date05/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-