| Model Number |
100613A |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The treatment to the patient was delayed.
|
| |
|
Event or Problem Description
|
|
Not nebulizing when connected to flowmeters.
|
| |
|
Additional Manufacturer Narrative
|
|
The treatment to the patient was delayed.In the absence of a sample or photos the defect could not be confirmed.However based on the customer information the most likely root cause was identified to be an occlusion within subassembly 0697.The ultimate risk of low identified during the risk analysis does not justify initiating a capa.Airlfie will continue to monitor product feedback and complaints through complaint trending analysis monthly for all products per our internal quality system procedures.
|
| |
|
Event or Problem Description
|
|
Not nebulizing when connected to flowmeters.
|
| |
|