| Catalog Number |
8065752201 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that the aspiration did not rise during cataract surgery (with intraocular lens (iol) implant).The surgery was successfully completed with the replacement of the fluidics management system (fms) pak.There was no patient harm.This report pertains to phacoemulsification tip involved in this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that the aspiration did not rise during phaco mode of cataract surgery (with intraocular lens (iol) implant).The surgery was successfully completed with the replacement of the fluidics management system (fms) pak.There was no patient harm.This report pertains to phacoemulsification tip involved in this reported event.Additional information received for this report the mode of the surgery was phaco mode.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A used fluidics management system (fms) in a tray was visually inspected and was tested using a machine console.The sample could be recognized by the console.The sample primed and tuned with a handpiece and a 0.9millimeter (mm) aspiration bypass system (abs) tip and infusion sleeve from lab stock, successfully and the service data could be retrieved.No system message code displayed on console screen.Cassette aspiration flow rate, and irrigation flow rate were measured and met specifications.No problem was found with the returned cassette fluidics.The root cause of the customer's complaint could not be established; the returned sample functioned per specification.This complaint has been reviewed and it is determined that no further actions will be pursed at this time as the sample met specifications.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).H10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|