| Catalog Number |
8065751658 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that an ophthalmic trocar was leaked during cataract-vitrectomy combinational surgery.The surgery was completed by replacing alternate trocar with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Sample was visually inspected and found to be nonconforming, damaged septum, hole was observed in septum.Leak testing could not be performed due to the damage of the sample.A review of the device history record traceable to the possible lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The exact root cause for this complaint is unknown; the damaged condition of the valves could have contributed to the reported event; how and when the valves became damaged cannot be determined from the evaluation performed.A contributing factor to the hole in sample is that the probe was actuated during insertion/removal of the probe from the trocar cannula during surgery.The exact root cause for this complaint is unknown therefore specific action for this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|