The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the segment fluid damage, the control body fluid damage, the lg connector fluid damage, the lg connector fluid damage, and the insertion flexible tube bump; however, these defects are not the main cause, and or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586 (image failure)"" and or the risk analysis results, it was evaluated to submit mdr.
|