| Model Number |
7760 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Insert (1316); Difficult to Advance (2920); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was fractured.A model-5f expo multipack 145 (5pkx3) was selected for percutaneous coronary intervention (pci).During the procedure, the expo catheter multipack found to have a stricture in the 'l' catheter of multiple catheters in the pack which made them difficult to flush and impossible to pass a wire.Furthermore, it was noted that the device was fractured.The device was completely removed from the patient's body and the procedure was completed with another of the same device.No complications were reported and the patient was stable after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: device codes: corrected from difficult to insert a150206 to difficult to advance a150205.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was fractured.A model-5f expo multipack 145 (5pkx3) was selected for percutaneous coronary intervention (pci).During the procedure, the expo catheter multipack found to have a stricture in the 'l' catheter of multiple catheters in the pack which made them difficult to flush and impossible to pass a wire.Furthermore, it was noted that the device was fractured.The device was completely removed from the patient's body and the procedure was completed with another of the same device.No complications were reported and the patient was stable after the procedure.
|
| |
|