This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: e3: reporter is a j&j sales representative.H4: the manufacture date is unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Udi: (b)(4).The lot number is unknown at this time.
|
|
It was reported by the sales rep that during shoulder arthroscopy labral repair procedure, bankart repair on (b)(6) 2024, using drill bit (stabilization) there were metal shavings that came off into the joint.It was reported that the shavings from drill bit were removed via shaver/suction.Another device was used to complete the procedure.There was a fifteen minute delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the investigation has been updated to reflect the correct information: investigation summary according to the information provided, it was reported that "while using drill bit (stabilization) there were metal shavings that came off into the joint.The shavings from drill bit were removed via shaver/suction".The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|