• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 HUMERAL HEAD TRIAL 42X14.5; SHOULDER INSTRUMENT - HEAD TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 HUMERAL HEAD TRIAL 42X14.5; SHOULDER INSTRUMENT - HEAD TRIAL Back to Search Results
Catalog Number 620042145
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tab broke.Surgery was not delayed due to the reported event.
 
Additional Manufacturer Narrative
Product complaint #:: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: tab broke.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that humeral head trial 42x14.5 has two internal tabs broken.Per inhance¿ shoulder system anatomic surgical technique (103832905 rev b) page 18, the humeral head trials are offered with a modular offset taper adapter trial.To assemble the two components, orient the offset taper adapter trial so that the taper and laser line are facing up.Align the laser mark on the offset taper adapter trial with the c and laser line on the underside of the humeral head trial so that the flexible nub on the offset taper adapter trial sits with the notch on the underside of the humeral head trial marked c.Then apply pressure to snap the two components together.Humeral head trial disassembly, page: 21: to remove the humeral head trial, assemble the head distractor to the impactor handle and place between the resection and the humeral head.To remove the offset taper adapter trial from the humeral head trial, rotate the offset taper adapter trial counterclockwise until the ramps unlock it from the head.From the e=eject position, additional counterclockwise rotation of the offset taper adapter trial will push it up the ramp and out of the humeral head trial.The observed breakage of the device was consistent with a component failure that may have been caused by exposure to unintended forces during the assembling or disassembling process.The overall complaint was confirmed as the observed condition of the humeral head trial 42x14.5 would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended user error.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: h4 corrected: h3, d4 the expiration date (12/31/2099) submitted in the medwatch #1 was reported in error.The device involved was an instrument and does not have an expiration date.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMERAL HEAD TRIAL 42X14.5
Common Device Name
SHOULDER INSTRUMENT - HEAD TRIAL
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18730776
Report Number1818910-2024-03664
Device Sequence Number17906271
Product Code HWT
UDI-Device Identifier10603295539179
UDI-Public10603295539179
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number620042145
Device Lot Number321043
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2024
Supplement Date Received by Manufacturer03/15/2024
04/09/2024
Initial Report FDA Received Date02/19/2024
Supplement Report FDA Received Date03/18/2024
04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-