| Catalog Number |
H938792 |
| Medical Device Problem Codes |
Disconnection (1171); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D4: lot # likely lot number: 60486689 (customer couldn't confirm the lot).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an exactamix valve set leaked.The port that connects to the empty bag, fully disconnected from the tubing resulting in leak of solution.This occurred under the hood during preparation.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot was manufactured between august 10, 2023 ¿ august 11, 2023.The device was received for evaluation.A visual inspection was performed, and it was noted that only the valve set silicone tubing was received and no valve body or connector was sent for evaluation which suggested the silicone outlet tubing was detached from the valve body outlet port.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|