| Catalog Number |
050-87215 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient¿s caretaker reported that during pd treatment there was a fluid leak encountered.There was an air detected in cassette warning during drain 4 of 6.Treatment was stopped and there was fluid found in the pump module and on the external cassette.In a follow up, the caregiver verified there were no symptoms, adverse events, or injuries requiring medical intervention required as a result of the reported event.The caregiver is letting the cycler air dry and will resume use.The cycler set is not available to return.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient¿s caretaker reported that during pd treatment there was a fluid leak encountered.There was an air detected in cassette warning during drain 4 of 6.Treatment was stopped and there was fluid found in the pump module and on the external cassette.In a follow up, the caregiver verified there were no symptoms, adverse events, or injuries requiring medical intervention required as a result of the reported event.The caregiver is letting the cycler air dry and will resume use.The cycler set is not available to return.
|
| |
|