Model Number 21987
Medical Device Problem Code Ejection Problem (4009)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Although requested, the device was not returned for evaluation.A device history record review could not be completed as a lot number was not provided.The trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on all available information, a root cause could not be conclusively determined.
 
Event or Problem Description
Reportedly, during iol implant procedure the injector piston overrode the lens.The incision was enlarged to remove the iol and a back up lens of an unknown model was implanted.A corneal suture was required.Additional information has been requested but not received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVISTA SIMPLIFEYE INJECTOR
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
BAUSCH + LOMB INC.
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
MEDICEL AG
dornierstrasse 11
altenrhein CH-94 23
SZ   CH-9423
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key18735495
Report Number0001313525-2024-70044
Device Sequence Number6234970
Product Code MSS
UDI-Device Identifier00757770547449
UDI-Public(01)00757770547449
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K192005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number21987
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENVISTA PRELOADED TORIC
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-