Catalog Number 3542630
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 01/02/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.  it is unknown at this time if the device will be made available for return.  as a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.  as such, the investigation will be closed.  if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation review could be performed.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.  this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.  this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.  if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent an unspecified breast surgery with unspecified mentor saline breast implants.Post-operatively, the patient experienced bilateral deflation of her prostheses, which was characterized by volume loss on both sides and diagnosed by a medical professional.At the time of this report, mentor has no information regarding explantation or an expected explantation date.The date of implantation was provided as an estimate.Follow-ups are being conducted.If more information becomes available, mentor will submit a supplemental report accordingly.This report is for implant b.Refer to manufacturing report number 1645337-2024-01978 for the contralateral event.
 
Additional Manufacturer Narrative
On march 18, 2024, mentor received additional information.The patient underwent removal surgery on (b)(6), 2024.Mentor received the device for evaluation.The complaint device was identified as a 225cc mentor siltex round moderate profile saline breast implant.Brand name: mentor siltex round moderate profile; lot: 5852730; catalog: 3542630.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.A discrepancy between the date of implantation and manufacturing date was observed.If more information becomes available, mentor will submit a supplemental report accordingly.On march 21, 2024, mentor completed a visual inspection and microscopic examination of the returned device.Device evaluation summary: visual analysis of the returned sample revealed that the breast implant was found to have a tear on the posterior view, measuring approximately 1.7 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have being damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key18735642
Report Number1645337-2024-01979
Device Sequence Number15025753
Product Code FWM
UDI-Device Identifier00081317001751
UDI-Public00081317001751
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2012
Device Catalogue Number3542630
Device Lot Number5852730
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2024
Supplement Date Received by Manufacturer03/18/2024
Initial Report FDA Received Date02/19/2024
Supplement Report FDA Received Date03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age45 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-