| Model Number |
N/A |
| Medical Device Problem Codes |
Device Alarm System (1012); Overheating of Device (1437); Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the unit is giving an overheat alarm and is smoking.The event timing is outside of surgery.There is no harm or delay reported.Due diligence is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the most recent repair record identified the following related repair: the exhaust tubing was crimped and was straightened.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
There is no additional information available.
|
| |
|