| Model Number |
OE-A63 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pentax medical america performed good faith effort to gather additional information regarding this event, however they are unable to obtain it at this time.If additional information becomes available, a supplemental report will be filed with the new information.Manufacturer mdr 9610877-2024-00013, duodenoscope model:ed34-i10t2, serial:unknown manufacturer mdr 9610877-2024-00014, dec distal cap model oe-a63 lot number (b)(6) manufacturer mdr 9610877-2024-00015, dec distal cap model oe-a63 lot number (b)(6).
|
| |
|
Event or Problem Description
|
|
Per the initial report, the first cap[dec] clicked on but detached before patient use, second cap[dec] clicked on but then detached before patient use, third cap[dec] attached and was used for a case and had difficulties removing from the endoscope.This event occurred at the time of during use.There was no report of patient harm.This event meets the requirements for fda reportability; however, submission of this report does not constitute an admission that medical personnel, user facility, importer, manufacturer or product caused or contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information g6: follow up #1.Evaluation summary: the facility did not respond to the gfe(good faith effort).Based on the investigation, since similar problems did not occur frequently at other facilities, we determined that it was a user error, but we were unable to determine the cause.A device history record(dhr) review was performed by the manufacturer.The dhr review confirmed the endoscope was manufactured on 24-oct-2022 under normal conditions, passed all required inspections, and was released accordingly.Also, there were no reworks or concessions and the dates of approval for shipment on 04-nov-2022 and actual date shipped were confirmed on 07-nov-2022.Pentax medical has not received any further information for this event and therefore, considers this medwatch report closed.
|
| |
|
Search Alerts/Recalls
|