MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; STERILE DISTAL END CAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; STERILE DISTAL END CAP Back to Search Results
Model Number OE-A63
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
Per the initial report, the first cap[dec] clicked on but detached before patient use, second cap[dec] clicked on but then detached before patient use, third cap[dec] attached and was used for a case and had difficulties removing from the endoscope.This event occurred at the time of during use.There was no report of patient harm.This event meets the requirements for fda reportability; however, submission of this report does not constitute an admission that medical personnel, user facility, importer, manufacturer or product caused or contributed to the event.
 
Additional Manufacturer Narrative
Pentax medical america performed good faith effort to gather additional information regarding this event, however they are unable to obtain it at this time.If additional information becomes available, a supplemental report will be filed with the new information.Manufacturer mdr 9610877-2024-00013, duodenoscope model:ed34-i10t2, serial:unknown, dec distal cap model oe-a63 lot number 0011112 manufacturer mdr 9610877-2024-00015, dec distal cap model oe-a63 lot number 0011112.
 
Additional Manufacturer Narrative
Evaluation summary the facility did not respond to the gfe(good faith effort).Based on the investigation, since similar problems did not occur frequently at other facilities, we determined that it was a user error, but we were unable to determine the cause.A device history record (dhr) review was performed by the manufacturer.The dhr review confirmed the endoscope was manufactured on 24-oct-2022 under normal conditions, passed all required inspections, and was released accordingly.Also, there were no reworks or concessions and the dates of approval for shipment on 04-nov-2022 and actual date shipped were confirmed on 07-nov-2022.Pentax medical has not received any further information for this event and therefore, considers this medwatch report closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
STERILE DISTAL END CAP
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 19600 12
JA  1960012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 19600 12
JA   1960012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key18735969
Report Number9610877-2024-00014
Device Sequence Number7243615
Product Code FDT
UDI-Device Identifier04961333226153
UDI-Public04961333226153
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K192245
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOE-A63
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2024
Supplement Date Received by Manufacturer01/26/2024
Initial Report FDA Received Date02/19/2024
Supplement Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-