Model Number 453944
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro mission drug-eluting stent system was selected for treatment of a lesion in the rca.The stent came off the balloon catheter in the guide.
 
Additional Manufacturer Narrative
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon of the delivery system is still well folded and shows no signs of inflation.Stent imprints are visible between the x-ray markers, indicating that the stent was initially crimped on the balloon.The stent is severely deformed at one end.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested by means of manufacturing process output monitoring.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Additional Manufacturer Narrative
Combination product: yes, corrected the initial reporter information.Reference capa# nc-24-004.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the balloon of the delivery system is still well folded and shows no signs of inflation.Stent imprints are visible between the x-ray markers, indicating that the stent was initially crimped on the balloon.The stent is severely deformed at one end.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested by means of manufacturing process output monitoring.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO MISSION (US) 2.5/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18737813
Report Number1028232-2024-00948
Device Sequence Number17952675
Product Code NIQ
UDI-Device Identifier07640130456006
UDI-Public(01)07640130456006(17)2509
Combination Product (Y/N)Y
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2025
Device Model Number453944
Device Catalogue Number453944
Device Lot Number08233112
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Supplement Date Received by Manufacturer05/02/2024
10/21/2024
Initial Report FDA Received Date02/20/2024
Supplement Report FDA Received Date05/07/2024
10/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;