| Catalog Number |
LSMU0800726 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a stent graft placement procedure, when the device was inserted into the sheath, it allegedly got caught and was difficult to be inserted it into the sheath.It was further reported that stent graft was allegedly dislodged.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 05/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.Therefore, a device history record review is not required.Investigation summary: the lifestream device was not returned for evaluation.No images and video files were provided for review.Therefore, the result of the investigation is inconclusive for the reported device incompatibility and stent dislodgment issues.The root cause for the reported device incompatibility and stent dislodgment issues could not be determined based upon the available information received from the field communications.Labeling review: the instructions for use for this lifestream device was reviewed and the following sections are applicable.B) indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.C) contraindications: the lifestream¿ balloon expandable vascular covered stent is contraindicated for use in: ¿ patients with uncorrected bleeding disorders.¿ patients who cannot receive recommended antiplatelet and/or anticoagulation therapy.¿ patients who are judged to have lesions that prevent complete expansion of the implant based on the lesion morphology, the diagnostic angiography and/or lesion response during pre-dilation.¿ lesions in which the lumen diameter post balloon angioplasty is insufficient for the passage of the endovascular system.¿ lesion locations subject to external compression.D) warnings: ¿ the lifestream¿ balloon expandable vascular covered stent is supplied sterile (by ethylene oxide) and is intended for single use only.Do not resterilize.After resterilization, the sterility of the product is not guaranteed because of an indeterminable degree of potential pyrogenic or microbial contamination which may lead to infectious complications.Cleaning, reprocessing and/or resterilization of the present medical device increases the probability that the device will malfunction due to potential adverse effects on components that are influenced by thermal and/or mechanical changes.¿ do not use the device if sterile packaging/pouch has been damaged or unintentionally opened prior to use.¿ use the device prior to the use by date specified on the package.¿ should excessive resistance be felt at any time during the procedure, do not force passage.¿ attempts to retract the covered stent into the sheath/guiding catheter may result in dislodgement and embolization of the covered stent.Maintain guidewire placement across the lesion and withdraw the endovascular system only until the proximal end of the covered stent is aligned with the tip of the sheath/guiding catheter.Do not attempt to remove an unexpanded covered stent through the sheath/guiding catheter.Remove the sheath/guiding catheter and endovascular system as a single unit.Attempting to remove an unexpanded covered stent by pulling it back into the sheath/guiding catheter may result in stent dislodgement.E) precautions: ¿ the device should only be used by physicians who are trained in endovascular procedures and are familiar with the required devices and materials, complications, side effects and hazards of peripheral vascular interventions.¿ anatomical variances and pre-existing conditions may complicate the procedure; always check the patient¿s medical history before starting the procedure and use caution when advancing the endovascular system through tortuous or difficult anatomy.¿ prior to device use, refer to the covered stent sizing table on the label and read the instructions for use.¿ keep dry.Keep away from sunlight.J) storage: store in a cool, dry place.Keep away from sunlight.Use the device prior to the use by date specified on the package.M) directions for use: site access and preparation: 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Air evacuation: 7.A 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.8.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.9.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.11.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.12.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent: 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.H10: d4 (expiration date: 05/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during the preparation of a stent graft placement procedure, when the device was inserted into the sheath, it allegedly got caught and was difficult to be inserted it into the sheath.It was further reported that the stent graft was allegedly dislodged.The procedure was completed by using another device.There was no patient contact.
|
| |
|