|
It was reported that during an ultrasound guided angioplasty procedure, when the pta balloon was removed, a small hole was allegedly found on the side near the tip of the balloon.The procedure was completed by using another device.There was no reported patient injury.
|
|
H10: as the lot number for the device was provided, a review of the device history record will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 04/2026).H11: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|