• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER INC SELF DRILL CANNULA DIA7.0MM X 80MM SHORT THREAD SCREW; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC SELF DRILL CANNULA DIA7.0MM X 80MM SHORT THREAD SCREW; SCREW, FIXATION, BONE Back to Search Results
Catalog Number MSD-010-70-080S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Arthritis (1723); Unspecified Infection (1930); Post Operative Wound Infection (2446)
Date of Event 01/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed. a review of the device history is not possible because the lot number was not communicated. should additional information become available, it will be provided on a supplemental report.H3 other text : device disposition unknown.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient underwent a device explant surgery due to infection caused by septic arthritis from the spine that traveled throughout the lower extremities of her body.The patient received a cement spacer.The infection was discovered during the purulent drainage from the lateral aspect of the ankle.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient underwent a device explant surgery due to infection caused by septic arthritis from the spine that traveled throughout the lower extremities of her body.The patient received a cement spacer.The infection was discovered during the purulent drainage from the lateral aspect of the ankle.
 
Additional Manufacturer Narrative
Correction - d1, h6 (results code and conclusion code).The reported event could be confirmed, based on available information and medical expert opinion.The device inspection was not possible as the product was not returned for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number was not communicated.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed: "implant removed due to hematogenic infection (bacteria from another infection site in the body carried by the bloodstream to the implant).This is a patient related issue, no a device related issue." based on investigation, the root cause was attributed to a patient related issue.The failure was caused by hematogenic infection.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF DRILL CANNULA DIA7.0MM X 80MM SHORT THREAD SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18738738
Report Number0001649390-2024-00055
Device Sequence Number17949170
Product Code HWC
UDI-Device Identifier00846832043071
UDI-Public00846832043071
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K131324
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberMSD-010-70-080S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/23/2024
Supplement Date Received by Manufacturer04/10/2024
Initial Report FDA Received Date02/20/2024
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
-
-