| Model Number |
UNIVERS VAULTLOCK GLENOID TRIAL, SMALL |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
On 2/1/2024, it was reported by a sales representative via (b)(4) that an ar-9236-01pp univers vaultlock glenoid trial central peg broke off.This was discovered during the removal of the trail during a procedure.The case was completed successfully, and the patient was not affected.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is confirmed.Functional testing was not performed due to one of the univers vaultlock¿ glenoid trial posts, small had broken off.No broken part was returned.Visual inspection identified one of the posts broke off and the edges around the device were chipped.The most likely cause can be attributed to use error due to the damage incurred during the insertion and/or removal process of the device.
|
| |
|