Model Number UNIVERS VAULTLOCK GLENOID TRIAL, SMALL
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
On 2/1/2024, it was reported by a sales representative via (b)(4) that an ar-9236-01pp univers vaultlock glenoid trial central peg broke off.This was discovered during the removal of the trail during a procedure.The case was completed successfully, and the patient was not affected.
 
Additional Manufacturer Narrative
Complaint is confirmed.Functional testing was not performed due to one of the univers vaultlock¿ glenoid trial posts, small had broken off.No broken part was returned.Visual inspection identified one of the posts broke off and the edges around the device were chipped.The most likely cause can be attributed to use error due to the damage incurred during the insertion and/or removal process of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS VAULTLOCK GLENOID TRIAL, SMALL
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18739268
Report Number1220246-2024-01036
Device Sequence Number6934444
Product Code KWS
UDI-Device Identifier00888867250673
UDI-Public00888867250673
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K161108
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUNIVERS VAULTLOCK GLENOID TRIAL, SMALL
Device Catalogue NumberAR-9236-01PP
Device Lot NumberS543GT000
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2024
Supplement Date Received by Manufacturer06/17/2024
Initial Report FDA Received Date02/20/2024
Supplement Report FDA Received Date12/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-