| Catalog Number |
9-ITV05F180/60 |
| Medical Device Problem Codes |
Retraction Problem (1536); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6)2024, a 5-4mm amplatzer duct occluder i was chosen for implantation utilizing a a 5f amplatzer torqvue delivery system.During preparation, the duct occluder had trouble advancing from the loader to the sheath.After multiple attempts to place the device, it was decided to recapture the occluder.When attempting to recapture the occluder into the delivery system, it would only recapture the waist of the device, but not the disc.The delivery system had to be removed while the occluder was still deployed at the end.A significant amount of time was added to the case but the patient remained hemodynamically stable throughout the procedure.There was no vessel or tissue damage.The patient remained stable.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 5-4mm amplatzer duct occluder i was chosen for implantation utilizing a a 5f a(mplatzer torqvue delivery system.During preparation, the duct occluder had trouble advancing from the loader to the sheath.After multiple attempts to place the device, it was decided to recapture the occluder.When attempting to recapture the occluder into the delivery system, it would only recapture the waist of the device, but not the disc.The delivery system had to be removed while the occluder was still deployed at the end.A significant amount of time was added to the case but the patient remained hemodynamically stable throughout the procedure.There was no vessel or tissue damage.The patient remained stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of difficult to advance, retraction problem, and positioning problem was reported.The device was received for investigation and no anomalies were found, with the device advancing and retracting a test valve under non-physiological conditions with ease.In addition, it was observed that the sheath was deformed and the loader was bent.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, a cause for the reported incidents could not be conclusively determined.The observe deformed sheath and kinked loader were due to post-procedural handling.There is no indication of a product quality issue with regards to manufacture, design, or labeling.H6 medical device problem code: code 2017 removed.
|
| |
|