• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TORQVUE DELIVERY SYSTEM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TORQVUE DELIVERY SYSTEM; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number 9-ITV05F180/60
Medical Device Problem Codes Retraction Problem (1536); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6)2024, a 5-4mm amplatzer duct occluder i was chosen for implantation utilizing a a 5f amplatzer torqvue delivery system.During preparation, the duct occluder had trouble advancing from the loader to the sheath.After multiple attempts to place the device, it was decided to recapture the occluder.When attempting to recapture the occluder into the delivery system, it would only recapture the waist of the device, but not the disc.The delivery system had to be removed while the occluder was still deployed at the end.A significant amount of time was added to the case but the patient remained hemodynamically stable throughout the procedure.There was no vessel or tissue damage.The patient remained stable.
 
Event or Problem Description
It was reported that on (b)(6) 2024, a 5-4mm amplatzer duct occluder i was chosen for implantation utilizing a a 5f a(mplatzer torqvue delivery system.During preparation, the duct occluder had trouble advancing from the loader to the sheath.After multiple attempts to place the device, it was decided to recapture the occluder.When attempting to recapture the occluder into the delivery system, it would only recapture the waist of the device, but not the disc.The delivery system had to be removed while the occluder was still deployed at the end.A significant amount of time was added to the case but the patient remained hemodynamically stable throughout the procedure.There was no vessel or tissue damage.The patient remained stable.
 
Additional Manufacturer Narrative
An event of difficult to advance, retraction problem, and positioning problem was reported.The device was received for investigation and no anomalies were found, with the device advancing and retracting a test valve under non-physiological conditions with ease.In addition, it was observed that the sheath was deformed and the loader was bent.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, a cause for the reported incidents could not be conclusively determined.The observe deformed sheath and kinked loader were due to post-procedural handling.There is no indication of a product quality issue with regards to manufacture, design, or labeling.H6 medical device problem code: code 2017 removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TORQVUE DELIVERY SYSTEM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18739905
Report Number2135147-2024-00748
Device Sequence Number20566362
Product Code DQY
UDI-Device Identifier00811806010755
UDI-Public(01)00811806010755(17)250731(10)8619043
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K072313
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Catalogue Number9-ITV05F180/60
Device Lot Number8619043
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2024
Supplement Date Received by Manufacturer04/17/2024
Initial Report FDA Received Date02/20/2024
Supplement Report FDA Received Date04/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age5 MO
Patient SexFemale
-
-