| Model Number |
N/A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported by customer, "midline was noticed to be broken off at the insertion site.The patient required x-rays to confirm the site of retained catheter fragments and surgically removed.Fragments measured 9cm." no other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.H3 other text : device not returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a break in the catheter was confirmed.The product returned for evaluation was one catheter hub with a small portion of catheter from an 18 ga powerglide pro.The investigation findings are consistent with catheter damage caused by contact with a sharp edged instrument such as a scalpel.The returned product sample was evaluated and a break was observed just distal of the luer.Microscopic examination of the catheter break confirmed it was typical of contact with a sharp bladed instrument, and the characteristics observed which supported this type of failure included: the fracture surface was reflective in nature and contained a striated pattern.This can occur due to pattern transfer of the sharpened edge typically found on a scalpel-type instrument.Sharply formed fracture edges planar shape to the fracture surface an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.H3 other text : evaluation findings are in section h.11.
|
| |
|
Event or Problem Description
|
|
It was reported via medwatch, "midline was noticed to be broken off at the insertion site.The patient required x-rays to confirm the site of retained catheter fragments and surgically removed.Fragments measured 9cm." no other information was provided.
|
| |
|