• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4009K
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the balloon of a large volume folfusor burst.This occurred before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4, h6, and h10: h4: the lot was manufactured between july 5, 2023 ¿ july 7, 2023.H10: the device was received for evaluation.A visual inspection was performed, and it was noted that there was fluid inside the bag that contained the unit.When the unit was removed from the bag, the cause of leak inside the bag was found to be an untightened red luer cap.The red luer cap is not a baxter product.The customer did not use the folfusor's blue winged luer cap; therefore, the cause of leak inside the bag was due to a use error.No evidence of rupture was observed from the bladder.The bladder was in normal condition.A functional leak test was performed, with no issues noted.Based on the sample analysis finding, the folfusor unit was determined to be conforming product.The reported condition of a ruptured bladder was not verified; however, a leak was discovered due to a user error.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18740939
MDR Text Key335813844
Report Number1416980-2024-00670
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4009K
Device Lot Number23G001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received03/13/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-