| Model Number |
WA22707S |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
As reported, the electrosurgical electrode loop was completely detached, fell off in the patient and was never retrieved.The issue occurred during therapeutic "rtuv" procedure.The intended procedure was completed.The device was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been 1 year since the subject device was manufactured.Based on the results of the investigation, the reported issue (distal end detachment/fall into patient) was likely due to wear and tear of the device.The loop at the distal end of the electrode may wear during use and can break, melt, or burn.This supplemental report includes a correction to b5 and g2 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported that no medical interventions were conducted to locate the broken device piece.It was confirmed that there was no patient harm.
|
| |
|
Search Alerts/Recalls
|