| Catalog Number |
9-AVP2-003 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 3mm amplatzer vascular plug 2 was chosen for implant.The plug was pushed out of the delivery system and the end was found to be deformed.There was no interaction with cardiac structures during deployment.There was no angulation or kink noticed in the delivery system.A replacement 3mm amplatzer vascular plug 2 was successfully implanted.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.The investigation confirmed the device met functional and visual specifications when analyzed at abbott.One image from field appeared to show the deformation of the device in bulbous formation which was not completely deployed from delivery cable.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|