| Catalog Number |
362760 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd received one (1) photo was provided for investigation.The photo was reviewed and the indicated failure mode for poor barrier separation was observed.Additionally, forty (40) retention samples from bd inventory were evaluated by visual examination and the issue of poor barrier separation was not observed.Complaints for sample quality are under statistical control for the month of january 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode poor barrier separation based on photos only.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® cpt¿ cell preparation tube with sodium citrate, there was poor barrier separation of the sample.Sample was recollected.There was no health impact or consequences reported.
|
| |
|