| Model Number |
VICM5_12.6 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint was reported for units within the same lot.H6: device history record (dhr) review: the dhr review indicated that the product has been manufactured within the established process parameters and that there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated a 12.6mm vicm5_12.6 implantable collamer lens, -06.00 diopter, tore/broke during loading and the lens was not implanted.The lens was replaced with another same model/length lens and the problem was resolved.The cause of the event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Originally the claim was determined to be reportable, but upon further review was determined there was no patient contact therefore not mdr reportable.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|