SYNTHES GMBH SPEEDTITAN COMPRESSION IMPL KIT 25X20X20; STAPLE, FIXATION, BONE Back to Search Results
Catalog Number SE-2520TI
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
Device report from germany reports an event as follows: it was reported that during a foot surgery on (b)(6) 2024, the placement instrument of two speed titan compression implant kits broke off.The doctor had to use forceps to insert the implants.The defective placement instruments were thrown away.No fragments remained in the patient.Procedure was completed successfully with no delay.There was no harm to patient or user.No additional surgical intervention planned.This report is for a speed titan compression impl kit 25x20x20.This is report 1 of 2 for (b)(4).Additional reports are captured under (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3,h4,h6 the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the sspeedtitan compression impl kit 25x20x20 was found broken from the implant inserter.The broken fragments were observed on the provided evidence.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed for speedtitan compression impl kit 25x20x20 there is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected data: h6: corrected photo investigation: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed a broken fragment however with the provided information is not possible to determine which pf the two devices involved on this complaint is broken.The broken fragments were observed on the provided evidence.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed however, it is unknown if the speedtitan compression impl kit 25x20x20 or speedtitan compression impl kit 18x18x15 is shown in the evidence.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDTITAN COMPRESSION IMPL KIT 25X20X20
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
BIO MEDICAL ENTERPRISES INC
14785 omicron dr # 205
san antonio TX 78245
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18743933
Report Number8030965-2024-02662
Device Sequence Number20428221
Product Code JDR
UDI-Device Identifier00810633021576
UDI-Public(01)00810633021576
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K142292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSE-2520TI
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2024
Supplement Date Received by Manufacturer03/14/2024
05/03/2024
Initial Report FDA Received Date02/20/2024
Supplement Report FDA Received Date03/29/2024
05/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SPEEDTITAN COMPRESSION IMPL KIT 18X18X15
Patient SexUnknown
-
-