ABBOTT MEDICAL SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Catalog Number TF-23A
Medical Device Problem Codes Device Stenosis (4066); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Aortic Valve Stenosis (1717); Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 10/23/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2015, a 23mm trifecta valve was implanted during an aortic valve replacement.On (b)(6) 2023, the valve was noted to have stenosis and regurgitation due to valve leaflet stiffness.The valve was suspected to have structural valve deterioration (svd).The valve was explanted and replaced with a 23mm epic supra valve.The patient was later discharged.
 
Additional Manufacturer Narrative
An event of stenosis and regurgitation due to valve leaflet stiffness was reported.It was reported that the valve was explanted due to suspicion of structural valve deterioration (svd).The investigation found circumferential fibrous pannus ingrowth with narrowing of inflow diameter.There was fibrous thickening on all leaflets.Calcifications present on leaflets 1 and 3.D.Leaflet 3 was torn with degenerative changes at the tear site.No acute inflammation was present.The device history record was also reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.The cause of the leaflet tear could not be conclusively determined; however, the degenerative changes noted to the tissue could have contributed to the tear formation.In addition, the fibrous pannus ingrowth noted had the potential to induce increased stress on adjacent leaflets and create an unbalanced stress relief distribution between all leaflets during coaptation, leading to leaflet tear and reduced durability.The calcifications noted could have contributed to the reported stenosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL LTDA. REG#3001883144
1301rua profvieira demendonça
bairro engenho nogueira 31.31 0-26
BR   31.310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18744170
Report Number2135147-2024-00776
Device Sequence Number15015134
Product Code LWR
UDI-Device Identifier05414734052030
UDI-Public(01)05414734052030(17)260420(10)4681816
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P100029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2026
Device Catalogue NumberTF-23A
Device Lot Number4681816
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/15/2024
Initial Report FDA Received Date02/20/2024
Supplement Report FDA Received Date04/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexMale
-
-