MEDOS INTERNATIONAL SARL GS 90, 95 CM, STRAIGHT, DISTAL; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number GS9095SD
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier and age were not provided.Section e.1: the initial reporter phone: (b)(6).Section h.9: fda correction/ removal reporting no.: 3007628272-02/02/2024-001-r.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (31154812) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the information available and without the complaint product available to be returned for analysis, the reported product issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint device; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during the procedure, the 95cm cerebase da guide sheath (gs9095sd / 31154812) broke.There was no patient harm nor significant delay reported.On 13-feb-2024, additional information was received.Per the information, the procedure was a cerebral thrombectomy targeting the left m1 segment of the middle cerebral artery (mca).The vessel was excessively tortuous with acute bends.This was not an adapt (direct aspiration first pass technique) case; the device was not snaked (advanced without wire / microcatheter).The cerebase did not become fractured into two separate pieces.The device was not replaced.The procedure was successfully completed.The information indicated that the concomitant embovac large bore catheter had ¿severe lengthening and deformation of the embovac during the retrieval maneuver due to kinking of the cerebase guiding catheter.¿ a continuous flush was maintained through the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GS 90, 95 CM, STRAIGHT, DISTAL
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key18744745
Report Number3008114965-2024-00212
Device Sequence Number12361307
Product Code QJP
UDI-Device Identifier10886704082323
UDI-Public10886704082323
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192804
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Catalogue NumberGS9095SD
Device Lot Number31154812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2024
Initial Report FDA Received Date02/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
IC 71, 132 CM, CE, ASP. IND.
Patient Age78 YR
Patient SexFemale
Patient Weight78 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-