The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the biopsy inlet t-piece residue inside biopsy t-piece aft tube.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the biopsy inlet t-piece.In addition, our technician confirmed that the insertion flexible tube buckled; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0588(channel)"" and/or the risk analysis results, it was evaluated to submit mdr.
|