The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the air tube clogged.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the air tube.In addition, our technician confirmed that the insertion flexible tube coating damage, the segment steel braid broken, the lcb (light carrying bundle) broken, the lg air supply tubes clogged, the air/water socket deformed, the operation channel (primary) cut, the remote control buttons cut, and the distal body worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0630(air/water & jet water channels)"" and/or the risk analysis results, it was evaluated to submit mdr.
|