C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT CATHETER; DIALYSIS CATHETER Back to Search Results
Catalog Number 5704150
Medical Device Problem Codes Fluid/Blood Leak (1250); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately two months post a dialysis catheter placement, the catheter allegedly had to be removed as a semi-emergency measure due to a leak.It was further reported that when water was injected through the branches of the removed catheter, a drop of water allegedly appeared at the point on the root of the red branch where the branches bifurcate.Furthermore, air bubbles were allegedly noted.Reportedly, the catheter was replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported leak and air in device issue, as no objective evidence was provided for review.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: g3, h6 (method).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately two months post a dialysis catheter placement, the catheter allegedly had to be removed as a semi-emergency measure due to a leak.It was further reported that when water was injected through the branches of the removed catheter, a drop of water allegedly appeared at the point on the root of the red branch where the branches bifurcate.Furthermore, air bubbles were allegedly noted.Reportedly, the catheter was replaced.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT CATHETER
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18746318
Report Number3006260740-2024-00685
Device Sequence Number8352219
Product Code MSD
UDI-Device Identifier00801741012907
UDI-Public(01)00801741012907
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5704150
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2024
Supplement Date Received by Manufacturer04/04/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age20 YR
Patient SexMale
Patient Weight99 KG
-
-