• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION Back to Search Results
Catalog Number 448100
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use with the bd phoenix¿ automated microbiology system, the instrument failed to correctly identify organisms.There was no report of patient impact.
 
Additional Manufacturer Narrative
E1.Initial reporter address: (b)(6).G5.Multiple 510(k): k020321, k020322, k020323, k040099, k131331.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H.6.Investigation summary: a failure for misidentification of results was reported on a phoenix 100 instrument (p/n 448100, s/n (b)(6).The customer reported that there were failures in the identification of microorganisms.The fse was dispatched for this case.Based on the work order summary, the fse proposed the equipment pre-installations must first be ensured before moving the equipment in the same counter near the bactec fx, place a film on half of the windows to prevent sun entry and install air conditioning.There is an electrical outlet in the laboratory with a neutral - ground voltage of 1.23 vac, phase - neutral 124.3 vac and phase - ground 125.1 vac and it is requested to verify the panel of that outlet in order to be able to draw a derivation for the equipment.Therefore, this is not confirmed failure of bd product.No samples were expected to be received as part of this complaint, and therefore no samples were returned, and no returned material investigation could occur.Dhr review is not required for this complaint as this complaint does not allege an early life failure or failure at installation and the configuration has changed since release from manufacturing due to service repairs/pms.Service history review was completed and revealed no further cases related to this failure mode.Bd quality will continue to monitor for trends associated with the failure mode described in this complaint.
 
Event or Problem Description
It was reported that during use with the bd phoenix¿ automated microbiology system, the instrument failed to correctly identify organisms.There was no report of patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ AUTOMATED MICROBIOLOGY SYSTEM
Common Device Name
SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18746799
Report Number1119779-2024-00128
Device Sequence Number15020501
Product Code LON
UDI-Device Identifier00382904481001
UDI-Public(01)00382904481001
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
SEE H10
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number448100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2024
Supplement Date Received by Manufacturer05/02/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/23/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-